pH
When a pH meter is reading incorrectly the simulator helps determine if the problem is related to the electrode or to the transmitter. It simulates pH and mV values: from 0-14 pH units in 1 pH unit increments (±414 mV). The simulator operates on a 9-volt battery and has a BNC connector, which attaches to the pH cable adapter.
Dissolved Oxygen
The simulator isolates the transmitter by mimicking the load of a DO sensor. The 4-pin connector attaches directly to the DO cable, in place of the sensor. In doing so, the simulator can help confirm whether fluctuations in the transmitter display are caused by the sensor, or caused by problems with the instrument or cable.
2
AM-9222
시뮬레이터
휴대형
DO Signal Simulator
DO 시뮬레이터
An essential diagnostic tool for every instrumentation lab. The simulator isolates the transmitter by mimicking the load of a DO sensor. In doing so, the simulator can help confirm whether fluctuations in the display are caused by the probe or EMF interferences. A two position switch allows simulation of either 0% or 100% saturation.
3
ECPHSIMULATOR
시뮬레이터
휴대형
PH Simulator
pH 시뮬레이터
Push button pH/mV selection
pH accuracy of ±0.05; mV range of ±2
USA/NIST pH buffer standards simulation
16 combination values
1 GΩ high input impedance test
Large display with bright LED
Splashproof keypad
Protective rubber boot doubles up as stand
Power-saving meter runs on 4 'AAA' x 1.4 V batteries