 특징 PH and DO Sensor Simulator
pH 시뮬레이터
DO 시뮬레이터 기능 pH
When a pH meter is reading incorrectly the simulator helps determine if the problem is related to the electrode or to the transmitter. It simulates pH and mV values: from 0-14 pH units in 1 pH unit increments (±414 mV). The simulator operates on a 9-volt battery and has a BNC connector, which attaches to the pH cable adapter.
Dissolved Oxygen
The simulator isolates the transmitter by mimicking the load of a DO sensor. The 4-pin connector attaches directly to the DO cable, in place of the sensor. In doing so, the simulator can help confirm whether fluctuations in the transmitter display are caused by the sensor, or caused by problems with the instrument or cable.
• pH Output: 0-14 pH in 1 pH increments • DO Output: 0% sat (0 nA), 100%
sat (~67 nA), 300% sat (~200 nA) • Power: 9-volt battery • 4-pin
connector for DO simulation • BNC jack for pH output • Auto Off
Feature • Tactile Membrane Keypad • Includes a 2-foot long pH cable to
accommodate the most common transmitter connections
AM-9504 관련 |
 |
|
|
궁금하신 점에 대해 해당 분야의 전문가가 상담해드립니다. 제품문의 버튼을 누르시면 입력폼이 출력됩니다. 담당자의 이름을 클릭하시면 직통 번호를 확인하실 수 있습니다. |
 |
|
|
|