| Wavelength  | 190~1,100nm  | 
| Technologh | Stabilized single-beam | 
| Source lamp | Xenon flashlamp | 
| Measuring mode | Concentrtion, absorbance,transmission,multi wavelengths,scans 
and kinetics in absorbance or transmission mode | 
|  Spectral bandwidth | 4nm | 
| Wavelength resolution | 1nm | 
| Wavelength accuracy |  ±1nm | 
| Photometric range | ±3.3A | 
| Absorbance resolution | 0.001A | 
| Absorbance accuracy | 0.003A bei<0.600A/0.5% of the reading  | 
|  Scan | Scans in 1nm increments with free selectable wavelenght range | 
| Cells | 16mm roung, 10, 20, 50mm rectangular with automatic cell size recognition | 
|  Test recognition | Automatic bar-code reading system for all Spectroquant® cell and reagent 
tests | 
| Method update | via internet/PC or via USB stick | 
|  Communication interfaces | 1 USB-Q, 1 USB-B, 1 RS232 | 
| Data storage | 1,000 single measured values; 4MB for scans and kinetics,i.e approx.100 scans(300~900nm) 
and 400 kinetics curves with 150 single values each | 
| Methods and profiles | Programmed methods of all Spectroquant® cell and reagent tests, 100 userdefined 
methods, 20 profiles for kinetic and absorption scans each | 
| Protection class | IP 31 | 
|  Power Supply | Country specific power supply,length of the connection cable: 2.0m | 
| Power requirements | 100-240V~/50-60Hz/0.70A | 
| Temperature | Operationg:+10°C to 35°C Storage:-25°C to 65°C | 
|  Dimensions |  404x197x314mm(widthxheightxdepth) | 
| Weight | approx. 4.5kg |