 특징 0~1000mg/l
(standard substance:kaolin)
4~20mA Analg
DC12V±10%  기능 calibration signal input
Analog (4-20mA) signal output: resistance load of 300Ω(Max.)
IP 68, maximum depth of 2 meters (underwater type)
| Model No |
TCS-1000 |
| Measuring range |
0-1000 (mg/l) (standard substance:kaolin) |
| Power supply |
DC12V±10% |
| Current draw |
At regular operation: 30mA (Max.), At cleaning operation: 240mA (Max.)
(excluding analog signal output) |
| Output |
Analog (4-20mA) signal output: resistance load of 300Ω(Max.) 4mA: 0 (mg/l)
20mA: 1000 (mg/l) Self-checking output: open collector (12VDC 20mA Max.)
|
| Input |
calibration signal input |
| Cleaning system |
Both sided swing wiper cleaning system |
| Time interval for
cleaning |
Clean once immediately after power-on, and subsequently clean once every 10
minutes. |
| Operating
temperature |
0-40°C (unfrozen) |
| Major material |
SUS 316L, sapphire glass, fluorocarbon rubber, EPDM, POM |
| Dimensions |
Ø 32x170mm (excluding horizontal hanging attachment) |
| Weight |
approx. 930g |
| Degree of Protection |
IP 68, maximum depth of 2 meters (underwater type) |
| Detector cable
length |
10m |
| Option |
Transmitter (SC-T4), Mounting attachment (TA-1), Maintenance kit (TC-MK) |
|