저울/기타계측기 >> 불소 >> 설치형 모델 : CA610
 
재고여부 : 45일 이내
 
특징
0.1~10mg/L fluoride


Accuracy: ± 10 % or ±0.10 mg/L whichever is greater

Air Purge: 0.1 cfm at 20 psig max. instrument quality air, 1/4-inch OD tubing,
quick-disconnect fitting,

Alarm: Two alarms, selectable for sample concentration, system warning or system shutdown. Each alarm equipped with an SPDT relay with contacts rated for 5 A resisitive load at 230 Vac.

Calibration Method: 5.0 mg/L fluoride standards

Compliance Certifications: CE approved, ETL listed to UL 1262, ETL certified to CSA 22.2 No. 142

Dimensions Metric (H x W X D): 34.3 mm x 41.9 mm x 19.1 mm

Display: LCD, 3 to 1/2 -digit measurement readout and six-character alphanumeric scrolling
text line

Drain Fitting: 1/2-inch ID flexible hose

Inlet: 1/4-inch OD polyethylene tubing, quick-disconnect fitting

Inlet Pressure: 1 to 10 psig (use sample restriction device
to reduce higher pressure samples)

Operating Humidity: 90 % at 40 °C (90% at 104°F) max.

Operating Temperature Range: 5 to 40 °C

Power Requirements (Amps): 2.5 A

Power Requirements (Hz): 50/60 Hz

Power Requirements (Voltage): 100 - 115/230 V AC

Power supply: selectable, 90 VA max. ,

Range: 0.1 to 10.0 mg/L Fluoride

Recorder Outputs: Una salida aislada del registrador, de 4 a 20 mA (puede ajustarse de 0 a 20 mA);

Sample Flow Rate: 200 to 500 ml/minute, (200 mL/min minimum required)

Sample Temperature: 5 to 40 °C

Storage Conditions: -40 to 60 °C (-40 to 140°F) without reagents
구성내역
NO.1
Fluoride Analyzer
  셋트일 경우엔 표준 부속품이 포함됨.